Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzer, ABL-700 series

Radiometer America, Inc / initiated 2005-06-17 / Terminated / root cause: software

Radiometer America Inc began a recall of Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzer, ABL-700 series on . The reason FDA records is "Software defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1317-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Radiometer America, Inc
Product
Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzer, ABL-700 series.
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
Software defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase.
Root cause tag
software
FDA root cause
Other
Action
The customers were apprised of this situation with a recall letter issued by the firm on 6/17/2005. In addition, new Operators Manuals were sent to the customers. In the letter, the customers were advised that a Service Engineer would be contacting them shortly in order to arrange for the installation of the software upgrade. The firm''s Service Engineers visited each customer location and installed software upgrades. Version 3.831 software was installed on the 700 Series Analyzers and Version 4.14 was installed on the 800 Series Analyzers. According to the firm, these software corrections were verified as having been completed by 3/10/2006.
Quantity in commerce
1635 devices
Distribution
Nationwide.
Product codes and lots
All analyzers manufactured from 1/1999-12/2005. The devices contain serial numbers 902-441RxxxNxxx. The Rxxx number denotes manufacturing run, and the Nxxx number denotes the individual instrument within the run.