ESP 140 Syringe Pump. Used for medication infusion. Model 140A and 140D, manufactured by Excelsior Medical...
Excelsior Medical Corp began a recall of ESP 140 Syringe Pump. Used for medication infusion. Model 140A and 140D, manufactured by Excelsior Medical Corporation, Neptune, NJ on . The reason FDA records is "Potential for alarm failure when the 140 pump completes infusion and/or the pump encounters a mid-infusion occlusion, when used in the 140 mode with a 140cc syringe. The ESP Pump works correctly when used in 60 mode...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0870-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Excelsior Medical LLC
- Product
- ESP 140 Syringe Pump. Used for medication infusion. Model 140A and 140D, manufactured by Excelsior Medical Corporation, Neptune, NJ.
- Product code
- FRN Pump, Infusion
- Reason for recall
- Potential for alarm failure when the 140 pump completes infusion and/or the pump encounters a mid-infusion occlusion, when used in the 140 mode with a 140cc syringe. The ESP Pump works correctly when used in 60 mode with a 60cc syringe or smaller.
- FDA root cause
- Other
- Action
- An 'Urgent Recall' letter was mailed certified return receipt mail on 3/22/06 to all customers per distribution list.
- Quantity in commerce
- 3911 units
- Distribution
- Nationwide and Quebec, Canada
- Product codes and lots
- All Serial Numbers since 1996.