Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzers, ABL-800 series, manufactured from 8/2004 through...
Radiometer America Inc began a recall of Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzers, ABL-800 series, manufactured from 8/2004 through 9/2005 on . The reason FDA records is "Under certain circumstances, the ABL 700 &800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1387-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Radiometer America, Inc
- Product
- Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzers, ABL-800 series, manufactured from 8/2004 through 9/2005.
- Product code
- JFP Electrode, Ion Specific, Calcium
- Reason for recall
- Under certain circumstances, the ABL 700 &800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and Sodium (Na).
- FDA root cause
- Other
- Action
- The consignees were notified of this recall by a letter, dated 9/29/2005. The letter informs the customers of this potential problem and the availability of new updated software which will correct the problem. The letter goes on to inform the customer that a Radiometer Service Engineer wil be contacting them shortly in order to arrange for a convenient time to install the software update. Along with the letter, the customer is provided a copy of an additional page and a CD concerning the function and operation of the new software, which is to be added to the customer''s Operator''s Manual.
- Quantity in commerce
- 196 devices
- Distribution
- Nationwide.
- Product codes and lots
- All analyzers manufactured from 8/2004-9/2005. The devices contain serial numbers 902-441RxxxNxxx. The Rxxx number denotes manufacturing run, and the Nxxx number denotes the individual instrument within the run.