VariSource AL 13199023 Implant Tube, Tongue/Soft Tissue, Puthawala/Syed:28cm Leader, 85cm Body, 15cm...
Alpha Omega Services Inc began a recall of VariSource AL 13199023 Implant Tube, Tongue/Soft Tissue, Puthawala/Syed:28cm Leader, 85cm Body, 15cm Draw-Down, Cable Insert, & Nylon Retaining Buttons: 20/PKG on . The reason FDA records is "There is a possibility that the diameter of the catheter does not meet specifications. As a result it may not be possible to correctly connect the catheter to a VariSource Quick Connect. This would prevent any...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0814-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alpha Omega Services Inc
- Product
- VariSource AL 13199023 Implant Tube, Tongue/Soft Tissue, Puthawala/Syed:28cm Leader, 85cm Body, 15cm Draw-Down, Cable Insert, & Nylon Retaining Buttons: 20/PKG
- Product code
- JAQ System, Applicator, Radionuclide, Remote-Controlled
- Reason for recall
- There is a possibility that the diameter of the catheter does not meet specifications. As a result it may not be possible to correctly connect the catheter to a VariSource Quick Connect. This would prevent any treatment. Furthermore, since the problem would not be apparent until after the catheter has been implanted an unnecessary surgical implantation procedure would have taken place.
- FDA root cause
- Other
- Action
- Recall Notification Letter: began February 22, 2006. Depth/Scope: 100% of the customer''s who purchased the affected products will be identified and contacted with a faxed and mailed Medical Device Recall notice and followed up with a phone call by the AOS Sales & Marketing personnel. All customers will be requested return all affected products, both used and unused.
- Quantity in commerce
- 93
- Distribution
- Nationwide and Argentina, Brasil, Canada, Israel, Korea, , Taiwan, United Kingdom & Venezuela
- Product codes and lots
- All Lots