Z-1120-06
Accu-Chek Ultraflex-1 infusion set; 10 mm 31'' (80 cm); Catalog number INF 04540883001 (Distributed U.S. and...
Disetronic Medical Systems / initiated 2006-03-31 / Terminated
Disetronic Medical Systems, Inc. began a recall of Accu-Chek Ultraflex-1 infusion set; 10 mm 31'' (80 cm); Catalog number INF 04540883001 (Distributed U.S. and Canada) and Ref. No 04626524001 (Distributed... on . The reason FDA records is "The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1120-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Disetronic Medical Systems
- Product
- Accu-Chek Ultraflex-1 infusion set; 10 mm 31'' (80 cm); Catalog number INF 04540883001 (Distributed U.S. and Canada) and Ref. No 04626524001 (Distributed outside of the the U.S. only).
- Product code
- LZG Pump, Infusion, Insulin
- Reason for recall
- The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- FDA root cause
- Other
- Action
- "U.S. and International distributors, patients and their physicians were notified via letters dated 3/31/06. Distributors were instructed to search for all part numbers affected by the recall and to provide all affected customers and sub-distributors with the recall notice. Customers were offered a replacement set or replacement with a different model. A press release was also issued on 4/3/06.
- Quantity in commerce
- 390,870 in US and Canada plus 10,510 outside N. America
- Distribution
- Worldwide including USA, Argentina, Austria, Barbados, Belgium, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Honduras, Hungary, Italy, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
- Product codes and lots
- Lots 31037UF, 31053UF, 31074UF, 31103UF, 3C130UF, 3D055UF, 3F060UF, 3F061UF, 3F095UF, 3G052UF, 3G127UF, 3I153UF, 3J068UF, 3J163UF, 3K047UF, 3K126UF, 3L144UF, 4B033UF, 4D044UF, 4E014UF, 4E141UF, 4F051UF, 4G035UF, 4G050UF, 4H159UF, 4H234UF, 4I056UF, 4I195UF, 4J076UF, 4K016UF, 4K164UF, 5C042UF, 5D211UF, 5E005UF, 5F074UF, 5H097UF, 5I056UF, 5I132UF, 5J149UF, 5K024UF, 5K176UF and 6B108UF.