Z-0797-06

SOMATOM Sensation CT Systems

Siemens Medical Solutions USA, Inc. / initiated 2006-03-17 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Sensation CT Systems on . The reason FDA records is "Firm became aware of a potential problem that could affect scan results and require patient exam to be repeated when using the customized ECG-gated scan protocols in the CT2006A and CT2006G software of their SOMATOM...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0797-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Sensation CT Systems
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Firm became aware of a potential problem that could affect scan results and require patient exam to be repeated when using the customized ECG-gated scan protocols in the CT2006A and CT2006G software of their SOMATOM Sensation CT Systems.
Root cause tag
software
FDA root cause
Other
Action
Firm has issued an advisory letter (3/28/2006) to affected customers per Update Instructions CT008/06/S. This letter informs customers about the potential problem and provides instructions to avoid this issue. A Siemens Service Engineer will visit affected sites to hand-deliver this letter and to delete any affected scan protocols, which can then be replaced with default protocols or with new customized protocols.
Quantity in commerce
86
Distribution
Nationwide. Product was distributed to medical facilities in AL, AR, AZ, CA, CO, DE, FL, GA, IA, IL IN, KS, KY, MA, MD, MI, MN, NC, NJ, NV, NY, OH, OK, PA, SC, TX, UT and WI.
Product codes and lots
Sensation 10 model number 7543015, Sensation 16 model number 7393114, Sensation 64 model number 8377520, and Sensation Open model number 8872017. Software versions CT2006A and CT2006G.