Z-1124-06

Accu-Chek Ultraflex-2 infusion set; 8 mm 43'' (110 cm); Catalog number INF 04540948001

Disetronic Medical Systems / initiated 2006-03-31 / Terminated

Disetronic Medical Systems, Inc. began a recall of Accu-Chek Ultraflex-2 infusion set; 8 mm 43'' (110 cm); Catalog number INF 04540948001 on . The reason FDA records is "The luer tube may break at the connection to the pump and, if this is not noticed by the patient, the patient may receive an under delivery of insulin". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1124-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Disetronic Medical Systems
Product
Accu-Chek Ultraflex-2 infusion set; 8 mm 43'' (110 cm); Catalog number INF 04540948001.
Product code
LZG Pump, Infusion, Insulin
Reason for recall
The luer tube may break at the connection to the pump and, if this is not noticed by the patient, the patient may receive an under delivery of insulin.
FDA root cause
Other
Action
"U.S. and International distributors, patients and their physicians were notified via letters dated 3/31/06. Distributors were instructed to search for all part numbers affected by the recall and to provide all affected customers and sub-distributors with the recall notice. Customers were offered a replacement set or replacement with a different model. A press release was also issued on 4/3/06.
Quantity in commerce
28,170 in US and Canada
Distribution
Worldwide including USA, Argentina, Austria, Barbados, Belgium, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Honduras, Hungary, Italy, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
Product codes and lots
Lots 5B066UF, 5C026UF, 5D203UF, 5E107UF, 5F016UF, 5G040UF, 5I027UF and 6B124UF.