Z-0824-06
Hobbs Medical 3 Prong Looped Retriever, a disposable device to be used by physician to remove foreign object...
Hobbs Medical, Inc. / initiated 2006-03-15 / Terminated / root cause: manufacturing process
Hobbs Medical, Inc. began a recall of Hobbs Medical 3 Prong Looped Retriever, a disposable device to be used by physician to remove foreign object or cauterize polyp specimens from the GI tract... on . The reason FDA records is "Diameter of device at the distal end would prevent it from passing through the endoscope's working channel". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0824-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hobbs Medical, Inc.
- Product
- Hobbs Medical 3 Prong Looped Retriever, a disposable device to be used by physician to remove foreign object or cauterize polyp specimens from the GI tract. REF: 5030 and 5030S
- Product code
- KOG
- Reason for recall
- Diameter of device at the distal end would prevent it from passing through the endoscope's working channel.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Hobbs Medical notfied customers by Registered Letter on 3/15/06, requesting return of recalled product.
- Quantity in commerce
- 254 units
- Distribution
- AK, IA, KY, MI, NH, NC,UT, WI, Canada, and UK
- Product codes and lots
- Lot Numbers: H11-05-040; H11-05-138; H11-05-139; H11-05-140;H11-05-141; H11-05-205.