Z-0899-06
Roche/Hitachi Glucose Hexokinase Reagent for use on the Hitachi Modular System; Catalog # 1929542...
Roche Diagnostics GmbH / initiated 2006-04-12 / Terminated
Roche Diagnostics Corp. began a recall of Roche/Hitachi Glucose Hexokinase Reagent for use on the Hitachi Modular System; Catalog # 1929542 (11929542216). In vitro diagnostic on . The reason FDA records is "A contaminant will change the pH of the R2 bottle, causing deterioration of the enzyme, and resulting in incorrect or no results being reported". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0899-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche/Hitachi Glucose Hexokinase Reagent for use on the Hitachi Modular System; Catalog # 1929542 (11929542216). In vitro diagnostic.
- Product code
- CFR Hexokinase, Glucose
- Reason for recall
- A contaminant will change the pH of the R2 bottle, causing deterioration of the enzyme, and resulting in incorrect or no results being reported.
- FDA root cause
- Other
- Action
- A recall notification letter dated 4/12/06 was sent to each customer instructing them to discontinue use of this lot of product.
- Quantity in commerce
- 269 kits
- Distribution
- Nationwide.
- Product codes and lots
- Lot 672652; exp. 6/30/2007.