FAST-FIX AB Curved Needle Delivery System, Used for meniscal repair. Product Number: 7209399

Smith & Nephew, Inc. / initiated 2006-03-24 / Terminated

Smith & Nephew, Inc. Endoscopy Division began a recall of FAST-FIX AB Curved Needle Delivery System, Used for meniscal repair. Product Number: 7209399 on . The reason FDA records is "The implant consists of a T1 and T2 anchor which are joined together by a suture. The T2 portion of the implant may not advance into the deployment position preventing deployment of the T2 anchor". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0798-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
FAST-FIX AB Curved Needle Delivery System, Used for meniscal repair. Product Number: 7209399
Product code
KGS Retention Device, Suture
Reason for recall
The implant consists of a T1 and T2 anchor which are joined together by a suture. The T2 portion of the implant may not advance into the deployment position preventing deployment of the T2 anchor.
FDA root cause
Other
Action
Smith & Nephew notified accounts by letter dated 3/24/06 issued by Federal Express. Users are requested to return product.
Quantity in commerce
1800 units
Distribution
Nationwide Foreign: Austria, Germany, France, Italy, Australia
Product codes and lots
Lot Numbers: 50141993, 50145519, 50146091,50146096, 50146362, 50148283, 50146365, 50153066, 50150381