Z-0798-06
FAST-FIX AB Curved Needle Delivery System, Used for meniscal repair. Product Number: 7209399
Smith & Nephew, Inc. / initiated 2006-03-24 / Terminated
Smith & Nephew, Inc. Endoscopy Division began a recall of FAST-FIX AB Curved Needle Delivery System, Used for meniscal repair. Product Number: 7209399 on . The reason FDA records is "The implant consists of a T1 and T2 anchor which are joined together by a suture. The T2 portion of the implant may not advance into the deployment position preventing deployment of the T2 anchor". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0798-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- FAST-FIX AB Curved Needle Delivery System, Used for meniscal repair. Product Number: 7209399
- Product code
- KGS Retention Device, Suture
- Reason for recall
- The implant consists of a T1 and T2 anchor which are joined together by a suture. The T2 portion of the implant may not advance into the deployment position preventing deployment of the T2 anchor.
- FDA root cause
- Other
- Action
- Smith & Nephew notified accounts by letter dated 3/24/06 issued by Federal Express. Users are requested to return product.
- Quantity in commerce
- 1800 units
- Distribution
- Nationwide Foreign: Austria, Germany, France, Italy, Australia
- Product codes and lots
- Lot Numbers: 50141993, 50145519, 50146091,50146096, 50146362, 50148283, 50146365, 50153066, 50150381