Medtronic Zinetics Manometric Catheter, Model 9012P2271. Water perfused, motility catheter. Distributed by...

Medtronic Gastroenterology / Urology / initiated 2006-04-06 / Terminated / root cause: labeling and UDI

Medtronic Gastroenterology / Urology began a recall of Medtronic Zinetics Manometric Catheter, Model 9012P2271. Water perfused, motility catheter. Distributed by Medtronic, Inc. 710 Medtronic Parkway NE... on . The reason FDA records is "Mislabeling- The 8 ports of the manometric catheter were incorrectly labeled 1,3,5,7,8,2,4,6 from the distal to the proximal end of the catheter. The correct port configuration is sequentially 1 through 8, from the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1159-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Gastroenterology / Urology
Product
Medtronic Zinetics Manometric Catheter, Model 9012P2271. Water perfused, motility catheter. Distributed by Medtronic, Inc. 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604 USA.
Product code
KLA Monitor, Esophageal Motility, Anorectal Motility, And Tube
Reason for recall
Mislabeling- The 8 ports of the manometric catheter were incorrectly labeled 1,3,5,7,8,2,4,6 from the distal to the proximal end of the catheter. The correct port configuration is sequentially 1 through 8, from the distal to the proximal end of the catheter.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Recall initiated to International accounts on 04/05/06 and within US on 04/06/06 via an Urgent Recall Notification. The letter informs them of the manufacturing defect, requests return for credit of any unused inventory and requests accounting of all previously used catheters. Distributors are instructed to sent a letter to their customers to request product return.
Quantity in commerce
94 catheters (US - 4; OUS - 90)
Distribution
Worldwide Distribution-including states of IL, NC, NY. and countries of Austria, Chile, Finland, Germany, Greece, Hong Kong, Italy, Japan, Lebanon, Norway, Peru, Portugal, Saudi Arabi, Spain
Product codes and lots
Lots 0001391 and 0001464