Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1104-06

AIRIS II, Altaire Nuclear Magnetic Resonance Imaging Device Systems

FUJIFILM Healthcare Americas Corporation / initiated 2006-04-28 / Terminated / root cause: software

Hitachi Medical Systems America Inc began a recall of AIRIS II, Altaire Nuclear Magnetic Resonance Imaging Device Systems on . The reason FDA records is "Software anomaly-A software defect can cause the slice line indicator to become mis-positioned during Multi-Planar Reconstructions (MPR) with the potential for patient mis-diagnosis". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1104-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
AIRIS II, Altaire Nuclear Magnetic Resonance Imaging Device Systems
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Software anomaly-A software defect can cause the slice line indicator to become mis-positioned during Multi-Planar Reconstructions (MPR) with the potential for patient mis-diagnosis.
Root cause tag
software
FDA root cause
Other
Action
The firm''s Field Service personnel will apprise the customers of this situation during scheduled visits at each customer location. During this visit, the corrected software will be installed by the Hitachi Field Service Technician. The installation of the software upgrades by the service technicians began on 4/28/2006. The firm anticipates that the correction of all units in their inventory as well as those units installed out-in-the-field will becompleted by 6/15/2006.
Quantity in commerce
5
Distribution
The devices were distributed to five (5) customers located in four (4)states: Colorado, Ohio, Nebraska and Texas
Product codes and lots
Serial Numbers: Altaire - L217, L219, L221, L224 and L225; AIRIS II - C686-C688* [*The AIRIS units remained in Hitachi's inventory and were not distributed].