Z-1445-06
Delphi IVantage Volumetric Ambulatory Infusion Pump/System; Ref nos. 12864051FRE, Rev. 8 and 12864051DEL...
Delphi Medical Systems / initiated 2006-03-29 / Terminated
Delphi Medical Systems began a recall of Delphi IVantage Volumetric Ambulatory Infusion Pump/System; Ref nos. 12864051FRE, Rev. 8 and 12864051DEL, Rev. 8 on . The reason FDA records is "Potential for under-infusion without alarm; cassette rollers stop moving , but the pump shaft continues rotating without alarm". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1445-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Delphi Medical Systems
- Product
- Delphi IVantage Volumetric Ambulatory Infusion Pump/System; Ref nos. 12864051FRE, Rev. 8 and 12864051DEL, Rev. 8.
- Product code
- FRN Pump, Infusion
- Reason for recall
- Potential for under-infusion without alarm; cassette rollers stop moving , but the pump shaft continues rotating without alarm.
- FDA root cause
- Other
- Action
- Distributors were notified via letter dated 3/29/05 instructing them to retrieve and return the product.
- Quantity in commerce
- 22 US and 147 international.
- Distribution
- Florida, Austria, Denmark, Finland, Germany, Norway, Sweden, Switzerland and United Kingdom.
- Product codes and lots
- All serial numbers.