Z-1445-06

Delphi IVantage Volumetric Ambulatory Infusion Pump/System; Ref nos. 12864051FRE, Rev. 8 and 12864051DEL...

Delphi Medical Systems / initiated 2006-03-29 / Terminated

Delphi Medical Systems began a recall of Delphi IVantage Volumetric Ambulatory Infusion Pump/System; Ref nos. 12864051FRE, Rev. 8 and 12864051DEL, Rev. 8 on . The reason FDA records is "Potential for under-infusion without alarm; cassette rollers stop moving , but the pump shaft continues rotating without alarm". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1445-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Delphi Medical Systems
Product
Delphi IVantage Volumetric Ambulatory Infusion Pump/System; Ref nos. 12864051FRE, Rev. 8 and 12864051DEL, Rev. 8.
Product code
FRN Pump, Infusion
Reason for recall
Potential for under-infusion without alarm; cassette rollers stop moving , but the pump shaft continues rotating without alarm.
FDA root cause
Other
Action
Distributors were notified via letter dated 3/29/05 instructing them to retrieve and return the product.
Quantity in commerce
22 US and 147 international.
Distribution
Florida, Austria, Denmark, Finland, Germany, Norway, Sweden, Switzerland and United Kingdom.
Product codes and lots
All serial numbers.