Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1197-06

Ossur Total Knee Junior, Model Number TK 1100, External Prosthetic knee

Ossur Americas / initiated 2006-03-29 / Terminated

Ossur North America Inc began a recall of Ossur Total Knee Junior, Model Number TK 1100, External Prosthetic knee on . The reason FDA records is "The company's initiation of the recall is based on a finding that some units of the Total Knee device may contain faulty pins based in the axis of the knee. Small cracks propagate through the pin cross section until the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1197-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ossur Americas
Product
Ossur Total Knee Junior, Model Number TK 1100, External Prosthetic knee.
Product code
ISY Joint, Knee, External Limb Component
Reason for recall
The company's initiation of the recall is based on a finding that some units of the Total Knee device may contain faulty pins based in the axis of the knee. Small cracks propagate through the pin cross section until the pin breaks, severely compromising the knee function.
FDA root cause
Other
Action
On 3/29/06 a press release was posted on OandP.com, an industry specific web site. An announcement listing involved serial numbers and contact phone numbers was also posted on the company''s web site, ossur.com. A two stage mailing has been constructed. The first letter sent to all customers March 30, 2006, describes the problem and models affected. The second letter, March 31, 2006, is customer specific and lists all of the involved serial numbers by model which were purchased by that customer. Telephone calls will be placed to customers who do not respond to the mailings.
Quantity in commerce
285
Distribution
Distribution worldwide in Austria, Canada, Columbia, Hong Kong, Japan, Korea, Kuwait, Malaysia, Mexico, Taiwan, & New Zealand. Involved states are: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, AND WY.
Product codes and lots
Serial numbers: 5787 ¿ 6207 including the following repaired units outside the stated range. 5339, 5471, 5478, 5670.