CALM software, version 02.09.01 This software is intended for use as a central monitoring system with signal...
LMS Medical Systems Ltd began a recall of CALM software, version 02.09.01 This software is intended for use as a central monitoring system with signal analysis and remote repeaters in the perinatal... on . The reason FDA records is "A deficiency in the software code can, in certain circumstances, cause data for the wrong patient to be displayed in the Tracing Review screen when a user is acknowledging an Alert". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0821-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- PeriGen Solutions Ltd.
- Product
- CALM software, version 02.09.01 This software is intended for use as a central monitoring system with signal analysis and remote repeaters in the perinatal clinical environment. It interfaces with standard fetal and maternal monitors and provides fetal surveillance, monitoring of labor progress, charting, reporting and data analysis, display and archiving functions.
- Product code
- HGM System, Monitoring, Perinatal
- Reason for recall
- A deficiency in the software code can, in certain circumstances, cause data for the wrong patient to be displayed in the Tracing Review screen when a user is acknowledging an Alert
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Consignees were first notified by telephone on 01/11/2005. A brief document was sent via email or fax on/about 01/13/2005.
- Quantity in commerce
- 8 units
- Distribution
- AL, IL, MD, NH, OH and Canada
- Product codes and lots
- Software version 02.09.01