Z-0821-06

CALM software, version 02.09.01 This software is intended for use as a central monitoring system with signal...

PeriGen Solutions Ltd. / initiated 2005-01-11 / Terminated / root cause: software

LMS Medical Systems Ltd began a recall of CALM software, version 02.09.01 This software is intended for use as a central monitoring system with signal analysis and remote repeaters in the perinatal... on . The reason FDA records is "A deficiency in the software code can, in certain circumstances, cause data for the wrong patient to be displayed in the Tracing Review screen when a user is acknowledging an Alert". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0821-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
PeriGen Solutions Ltd.
Product
CALM software, version 02.09.01 This software is intended for use as a central monitoring system with signal analysis and remote repeaters in the perinatal clinical environment. It interfaces with standard fetal and maternal monitors and provides fetal surveillance, monitoring of labor progress, charting, reporting and data analysis, display and archiving functions.
Product code
HGM System, Monitoring, Perinatal
Reason for recall
A deficiency in the software code can, in certain circumstances, cause data for the wrong patient to be displayed in the Tracing Review screen when a user is acknowledging an Alert
Root cause tag
software
FDA root cause
Other
Action
Consignees were first notified by telephone on 01/11/2005. A brief document was sent via email or fax on/about 01/13/2005.
Quantity in commerce
8 units
Distribution
AL, IL, MD, NH, OH and Canada
Product codes and lots
Software version 02.09.01