Medtronic Paceart 2004 Second Edition & Paceart 2005 First Edition versions of the Paceart Gateway Software...
Medtronic Inc. Cardiac Rhythm Managment began a recall of Medtronic Paceart 2004 Second Edition & Paceart 2005 First Edition versions of the Paceart Gateway Software. The Medtronic Paceart System is a personal... on . The reason FDA records is "The Paceart System could inappropriately insert data from a patient's Medtronic CareLink transmission into another patient's Paceart record. An issue associated with the automated batch transfer of implanted cardiac...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0911-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic Paceart 2004 Second Edition & Paceart 2005 First Edition versions of the Paceart Gateway Software. The Medtronic Paceart System is a personal computer based pacemaker-testing system, a 12-lead electrocardiograph, and a transtelephonic receiving station. It provides data that can be used to analyze implanted pacemaker performance based on electrocardiographic measurements, either taken directly from the patient or programmer. The system can measure, store, and display any of the 12 standard leads. Reports and charts are available by means of a laser printer. The system also includes a database that collects and stores patient data. Optional software allows the device to function as an unattended transtelephonic receiving station.
- Product code
- DPS Electrocardiograph
- Reason for recall
- The Paceart System could inappropriately insert data from a patient's Medtronic CareLink transmission into another patient's Paceart record. An issue associated with the automated batch transfer of implanted cardiac device data from the Medtronic CareLink network to the Paceart System exists.
- FDA root cause
- Other
- Action
- Affected user and potential users of the Gateway product were notified via phone and/ or a customer notification. The notification offers interim options until a software update is available. Customers are asked to respond back to Metronic with an enclosed Acknowledgement form.
- Quantity in commerce
- 210
- Distribution
- Nationwide
- Product codes and lots
- Paceart 2004 second edition and Paceart 2005 first edition