Z-0827-06

Guidant CONTAK RENEWAL 4 AVT (model M170). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile...

Boston Scientific Corporation / initiated 2006-04-07 / Terminated

Guidant Corporation began a recall of Guidant CONTAK RENEWAL 4 AVT (model M170). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Hamline Avenue North... on . The reason FDA records is "77 distributed devices may have a cracked layer of insulation in a flexible hybrid circuit, leading to internal shorting during charging such as during routine capacitor formation or for defibrillation. This would...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0827-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Guidant CONTAK RENEWAL 4 AVT (model M170). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Cardiac resynchronization therapy defibrillators (CRT-D) provide both atrial and ventricular tachyarrhythmia and cardiac resynchronization therapies. Ventricular tachyarrhythmia therapy is for the treatment of ventricular tachycardia (VT) and ventricular fibrillation (VF), rhythms that are associated with sudden cardiac death (SCD). Atrial tachyarrhythmia therapy is for the treatment of supraventricular tachcardia (SVT) and atrial fibrillation (AF). Cardiac resynchronization therapy is for the treatment of heart failure (HF) and uses biventricular electrical stimulation to synchronize ventricular contractions. This device also uses accelerometer-based adaptive-rate bradycardia therapy. Cardioversion/defibrillation therapies include a range of low-and high-energy shocks using either a biphasic or monophasic waveform.
Product code
NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Reason for recall
77 distributed devices may have a cracked layer of insulation in a flexible hybrid circuit, leading to internal shorting during charging such as during routine capacitor formation or for defibrillation. This would likely render the device non-functional, resulting in no output and/or telemetry communications.
FDA root cause
Other
Action
Guidant Sales Representatives are retrieving non-implanted devices April 10-12, 2006. If a device is implanted the physician is sent an Urgent Medical Device Safety Information & Corrective Action letter, April 11-14, 2006. This letter describes the issue, states that Guidant is retrieving all non-implanted devices and that Guidant is currently conducting tests on retrieved devices. Physicians will be advised once more information is available.
Quantity in commerce
20 devices
Distribution
Nationwide. AZ, FL, GA, IL, IN, KY, MA, ME, MI, MN, NC, NE, NY, OH, PA, TN, TX, UT, VA, WI. OUS to include: Australia, Canada, Germany, Italy, Netherlands, Portugal, Slovenia, Spain.
Product codes and lots
model M170 serial numbers: 200099, 200103, 200104.