Locator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing...

Abbott Medical / initiated 2006-03-29 / Terminated / root cause: packaging

Smiths Medical MD, Inc. began a recall of Locator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing and identified as PORT-A-CATH II Fluoro-Free, Venous... on . The reason FDA records is "One lot of Locator-Wand Cover sterile pouches may have an insufficient seal and product sterility could be compromised. The Locator-Wand Cover is supplied with the PORT-A-CATH II AND P.A.S. PORT systems. The packaged...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0801-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Locator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing and identified as PORT-A-CATH II Fluoro-Free, Venous Access System, PolyFlow Polyurethane Catheter, 2.6mm O.D. tray. REF 21-4655-24. Contents: Portal with ULTRA-LOCK Connector, Radiopaque PolyFlow Polyurethane Catheter with CATH-FINDER Sensor Wire, Locator-Wand cover, Nonvented Luer Cap, 20G PORT-A-CATH needle, 16G Blunt Needle, 20G GRIPPER PLUS Non-Coring Safety needle, 18G Extra Thin Wall Introducer Needle, 8.5 Fr Introducer, ''J'' Guidewire, Syringes, Vein Pick, Tunneling Tool, 20G 90-degree PORT-A-CATH needle, 3 Point-Lok Sharps Safety Devices, Product Information. EtO Sterilized. Single Use. Deltec, Inc., St. Paul, MN 55112 U.S.A.
Product code
MMP Cover, Barrier, Protective
Reason for recall
One lot of Locator-Wand Cover sterile pouches may have an insufficient seal and product sterility could be compromised. The Locator-Wand Cover is supplied with the PORT-A-CATH II AND P.A.S. PORT systems. The packaged Wand-Cover is placed on the outside of the PORT-A-CATH II AND P.A.S. PORT tray.
Root cause tag
packaging
FDA root cause
Other
Action
An Urgent Product Safety & Recall Notification, dated 03/29/06, was mailed to domestic customers and emailed to international customers. The letter described the issue, identified re-order number and lot numbers affected and gave 3 options. The options include to use only the sterile tray and discard the wand cover from the shelf box; remove and discard Locator-Wand Cover pouches and request replacement Locator-Wand Covers; or return the entire device for a credit or replacement.
Quantity in commerce
73 trays
Distribution
Nationwide. AR, CA, DE, FL, IL, KY, LA, MA, MD, MI, MN, MS, NJ, NY, OK, PA, TN, TX, VA, WI. Germany, Italy, Netherlands, Singapore, Spain
Product codes and lots
Re-order Number 21-4655-24, Lot numbers M42087, M43079