Z-0834-06

Newton IQ 4 Lead Cycler Set with stay-safe patient connectors, used for acute and chronic peritoneal...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2006-03-22 / Terminated

Fresenius Medical Care North America began a recall of Newton IQ 4 Lead Cycler Set with stay-safe patient connectors, used for acute and chronic peritoneal dialysis. Catalog Number: 050-87028 on . The reason FDA records is "Stay safe trigger body may not connect to the extension set and result in associated peritonitis". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0834-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Newton IQ 4 Lead Cycler Set with stay-safe patient connectors, used for acute and chronic peritoneal dialysis. Catalog Number: 050-87028
Product code
FKX System, Peritoneal, Automatic Delivery
Reason for recall
Stay safe trigger body may not connect to the extension set and result in associated peritonitis
FDA root cause
Other
Action
Fresenius Medical notified customers by telephone on 3/24/06 for initial lots recalled and followed by letter dated 3/30/06 to include all lots numbers recalled.
Quantity in commerce
141,770 units
Distribution
Nationwide Canada
Product codes and lots
Lot Numbers: 5SR119; 52R124; 5SR125; 5SR149; 5SR150; 5SR173; 6AR014; 6AR015; 6AR032; 6AR044; 6AR045; 6AR055; 6AR069; 6AR085; 06AR086; 6AR267; 6AR280;6AR298; 6BR013;6BR031; 6BR040; 6BR043; 6BR064; 6BR075; 6BR091; 6BR092; 6BR110; 6BR127; 6BR141; 6BR281