EP-4 Computerized Stimulator -- 2 channel and 4 channel. The product consists of a touchscreen portable PC...
EP Medsystems began a recall of EP-4 Computerized Stimulator -- 2 channel and 4 channel. The product consists of a touchscreen portable PC controlling a two channel dedicated EP stimulator... on . The reason FDA records is "Under certain circumstances the device may exhibit the following behaviors: 1)-Initiation of a rapid stimulation train without command, when the device is in use in conjuction with an electrosurgical unit or RF ablation...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0914-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- EP-4 Computerized Stimulator -- 2 channel and 4 channel. The product consists of a touchscreen portable PC controlling a two channel dedicated EP stimulator. The stimulator is capable of single channel, simultaneous and sequential stimulation at programmed output settings using built-in protocols, and includes an emergency stimulation function.
- Product code
- JOQ Generator, Pulse, Pacemaker, External Programmable (For Electrophysiological Studies Only)
- Reason for recall
- Under certain circumstances the device may exhibit the following behaviors: 1)-Initiation of a rapid stimulation train without command, when the device is in use in conjuction with an electrosurgical unit or RF ablation unit-2)-Presence of a DC offset voltage on an output channel-3)-Inability to synchronize with a non-EP MedSystems recording system
- FDA root cause
- Other
- Action
- The firm initiated the recall on April 4, 2006 through a notification letter sent to the electrophysiology lab managers at the healthcare institutions and two domestic distributors to which the device has been distributed. The firm used a traceable courier service to deliver notification of the recall.
- Quantity in commerce
- 187 units
- Distribution
- Product was shipped to hospitals/users throughout the US. The firm has a Veterans Affair contract number V797P-4650A. Also the firm has shipped the devices to Germany, United Kingdom, Netherlands, Italy, Japan, France, Saudi Arabia, Hong Kong, Syria, Malaysia, Sweden, Armenia, Australia, Canada, Iceland, S. Africa, and Taiwan.
- Product codes and lots
- Model number: EP-4-02 and EP-4-04. EP40312-001 through 010; EP40403-001 through 014; EP40406-001 through 042; EP40408-001 through 039; EP40410-001 through 040; EP40411-001 thourgh 040; EP40504-001 through 040; EP40506-001 through 040; EP40508-001 through 020; EP40510-001 through 020; EP40511-001 through 040.