Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0913-06

Bard TruGuide Coxail Biopsy Needle with Depth Stop and additional Blunt Tip Stylet, Catalog Number C1816B...

Bard Peripheral Vascular Inc / initiated 2006-05-01 / Terminated / root cause: manufacturing process

Bard Peripheral Vascular Inc began a recall of Bard TruGuide Coxail Biopsy Needle with Depth Stop and additional Blunt Tip Stylet, Catalog Number C1816B, Device Listing Number E148687. The Bard TruGuide... on . The reason FDA records is "Manufacturing change that has introduced variability in the length of the inner stylet. When the inner stylet length is inserted into the cannula component, the exposed stylet length may be up to 8mm longer than expected". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0913-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Peripheral Vascular Inc
Product
Bard TruGuide Coxail Biopsy Needle with Depth Stop and additional Blunt Tip Stylet, Catalog Number C1816B, Device Listing Number E148687. The Bard TruGuide Coaxial Biopsy Needle is a three-part device consisting of an outer cannula with an attached female luer lock hub, and inner stylet with an attached male luer lock hub, and a flexible slip rink depth stop.
Product code
FCG Biopsy Needle
Reason for recall
Manufacturing change that has introduced variability in the length of the inner stylet. When the inner stylet length is inserted into the cannula component, the exposed stylet length may be up to 8mm longer than expected.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Recall letters were sent to all consignees (Radiology Department Manager with a copy to the Hospital Administrator) via Federal Express overnight delivery with signature receipt required on May 1, 2006. Each letter included a Recall and Effectiveness Check Form for each consignee to fill out and fax to Bard Peripheral Vascular, Inc (BPV).
Quantity in commerce
225
Distribution
The product was distributed to the following five states: 1) California, 2) Florida 3) New Jersey, 3) Missouri, 4) Texas, 5) Washington. The product was also distributed to Canada and Latin America. Note: units distributed to Latin America remained in Bard''s distribution system and have been returned.
Product codes and lots
Lot # REP JO196