Z-0915-06
Endopath Probe Plus II, Hook Electrode, 5 mm Shaft, 34 cm length, hollow tip electrdode, Product Code EPS01
Ethicon Endo-Surgery Inc / initiated 2006-04-20 / Terminated / root cause: packaging
Ethicon Endo-Surgery Inc began a recall of Endopath Probe Plus II, Hook Electrode, 5 mm Shaft, 34 cm length, hollow tip electrdode, Product Code EPS01 on . The reason FDA records is "There is the possibility of a seal void in the Probe Plus packaging which may compromised the sterility of the device". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0915-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon Endo-Surgery Inc
- Product
- Endopath Probe Plus II, Hook Electrode, 5 mm Shaft, 34 cm length, hollow tip electrdode, Product Code EPS01.
- Product code
- GCJ Laparoscope, General & Plastic Surgery
- Reason for recall
- There is the possibility of a seal void in the Probe Plus packaging which may compromised the sterility of the device.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- The recalling firm sent a recall letter by United Parcel Services to all consignees for delivery to each consignee on 4/24/06. The letter requested subaccount address return information to the recalling firm.
- Quantity in commerce
- 414 cases/6 devices per case.
- Distribution
- The recalled product was distributed throughout the United States and to foreign consignees including Argentina, Australia, Austria, Bahrain, Brazil, Canada, China, Colombia, Czech Republic, Dominican Republic, Ecuador, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Korea, Lebanon, Malaysia, Mexico, Netherlands, New Zealand, Pakistan, Peru, Philippines, Poland, Portugal, Puerto Rico, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uruguay, Venezuela.
- Product codes and lots
- Lots C4D111, C4DD09, C4DD0H, C4DE5L, C4DK67, C4DM01, C4DM03, and C4DN03.