Z-0886-06
SYNCHRON Systems Vancomycin Reagent, in vitro diagnostic. Part Number 474824
Siemens Healthcare Diagnostics Inc. / initiated 2006-03-30 / Terminated
Beckman Coulter Inc began a recall of SYNCHRON Systems Vancomycin Reagent, in vitro diagnostic. Part Number 474824 on . The reason FDA records is "Beckman Coulter has received reports of patient samples and quality control recovery shifting low when using Vancomycin reagent lots M507219 and M511327. Internal QC recovery was as much as 15% lower than the assigned...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0886-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- SYNCHRON Systems Vancomycin Reagent, in vitro diagnostic. Part Number 474824.
- Product code
- LEH Radioimmunoassay, Vancomycin
- Reason for recall
- Beckman Coulter has received reports of patient samples and quality control recovery shifting low when using Vancomycin reagent lots M507219 and M511327. Internal QC recovery was as much as 15% lower than the assigned target value.
- FDA root cause
- Other
- Action
- A Product Corrective Action (PCA) letter was be sent the week of MAR 30, 2006 to all SYNCHRON Systems Vancomycin Reagent customers of lot numbers M507219 and M511327 informing them to discontinue use and discard reagent. (Letters were sent by US mail)
- Quantity in commerce
- 3147
- Distribution
- Nationwide & Canada
- Product codes and lots
- Lot numbers M507219 & M511327