Z-0886-06

SYNCHRON Systems Vancomycin Reagent, in vitro diagnostic. Part Number 474824

Siemens Healthcare Diagnostics Inc. / initiated 2006-03-30 / Terminated

Beckman Coulter Inc began a recall of SYNCHRON Systems Vancomycin Reagent, in vitro diagnostic. Part Number 474824 on . The reason FDA records is "Beckman Coulter has received reports of patient samples and quality control recovery shifting low when using Vancomycin reagent lots M507219 and M511327. Internal QC recovery was as much as 15% lower than the assigned...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0886-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
SYNCHRON Systems Vancomycin Reagent, in vitro diagnostic. Part Number 474824.
Product code
LEH Radioimmunoassay, Vancomycin
Reason for recall
Beckman Coulter has received reports of patient samples and quality control recovery shifting low when using Vancomycin reagent lots M507219 and M511327. Internal QC recovery was as much as 15% lower than the assigned target value.
FDA root cause
Other
Action
A Product Corrective Action (PCA) letter was be sent the week of MAR 30, 2006 to all SYNCHRON Systems Vancomycin Reagent customers of lot numbers M507219 and M511327 informing them to discontinue use and discard reagent. (Letters were sent by US mail)
Quantity in commerce
3147
Distribution
Nationwide & Canada
Product codes and lots
Lot numbers M507219 & M511327