Z-1357-06

ACCU-CHEK Advantage Blood Glucose Meter, Cat. No. 3332586; Ref. no. 03332586001 with meter model no. 2138930

Roche Diagnostics GmbH / initiated 2006-06-22 / Terminated

Roche Diagnostics Corp. began a recall of ACCU-CHEK Advantage Blood Glucose Meter, Cat. No. 3332586; Ref. no. 03332586001 with meter model no. 2138930 on . The reason FDA records is "The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that the test strip may be damaged or test was...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1357-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
ACCU-CHEK Advantage Blood Glucose Meter, Cat. No. 3332586; Ref. no. 03332586001 with meter model no. 2138930.
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that the test strip may be damaged or test was not performed correctly.
FDA root cause
Other
Action
A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.
Quantity in commerce
55,189 meters (for all Recalls Z-1354-1371-06)
Distribution
Worldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.
Product codes and lots
All units.