Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1358-06

ACCU-CHEK Comfort Curve, Cat. No. 2139634. Ref. numbers 03856348001 and 12139642001 both with meter model...

Roche Diagnostics GmbH / initiated 2006-06-22 / Terminated

Roche Diagnostics Corp. began a recall of ACCU-CHEK Comfort Curve, Cat. No. 2139634. Ref. numbers 03856348001 and 12139642001 both with meter model number 2138930 on . The reason FDA records is "The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a strip error". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1358-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
ACCU-CHEK Comfort Curve, Cat. No. 2139634. Ref. numbers 03856348001 and 12139642001 both with meter model number 2138930.
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a strip error.
FDA root cause
Other
Action
A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.
Quantity in commerce
55,189 meters (for all Recalls Z-1354-1371-06)
Distribution
Worldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.
Product codes and lots
All units.