Coated Vicryl (polyglactin 910) suture. Needle Type CTXB Suture size: 1. Vicryl suture is a synthetic...

Ethicon, Inc / initiated 2006-05-04 / Terminated / root cause: packaging

Ethicon, Inc. began a recall of Coated Vicryl (polyglactin 910) suture. Needle Type CTXB Suture size: 1. Vicryl suture is a synthetic absorbable sterile surgical suture comprised of a... on . The reason FDA records is "Misalignment of packaging sealing platform created open seals and/or open channels in the packaging seal causing degradation of the sutures". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1087-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon, Inc
Product
Coated Vicryl (polyglactin 910) suture. Needle Type CTXB Suture size: 1. Vicryl suture is a synthetic absorbable sterile surgical suture comprised of a copolymer made from 90% gycolide and 10%L-lactide.
Product code
GAM Suture, Absorbable, Synthetic, Polyglycolic Acid
Reason for recall
Misalignment of packaging sealing platform created open seals and/or open channels in the packaging seal causing degradation of the sutures.
Root cause tag
packaging
FDA root cause
Other
Action
The firm sent out recall notification on 5/4/2006. The US consignees were sent a letter via return receipt.
Quantity in commerce
1157 dozens
Distribution
Nationwide, Japan, Canada, and Taiwan.
Product codes and lots
Product Codes/Product Lot Numbers-Expiration Dates: J774D/SB2371-01/2008; JB725/SB2416-01/2008; J864D/SB2426-01/2008.