Talon Straight, 25 cm--Electrosurgical Device--RITA * Medical Systems, Inc. * StarBurst Talon * * REF...
Rita Medical Systems, Inc. began a recall of Talon Straight, 25 cm--Electrosurgical Device--RITA * Medical Systems, Inc. * StarBurst Talon * * REF: 700-102846 * 4 cm Diameter * 25 cm Length * Sterile/R *... on . The reason FDA records is "Misbranding; Product may have the wrong device handle which incorrectly identifies the product and also has incorrect deployment depth marker". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0887-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Talon Straight, 25 cm--Electrosurgical Device--RITA * Medical Systems, Inc. * StarBurst Talon * * REF: 700-102846 * 4 cm Diameter * 25 cm Length * Sterile/R * Catalog # 700-102846-RITA * Manufactured by: Rita Medical Systems, Inc., One Horizon Way, Manchester, Georgia 31816, USA
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Misbranding; Product may have the wrong device handle which incorrectly identifies the product and also has incorrect deployment depth marker.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Consignees were first notified by telephone on 05/02/2006. A follow up letter was sent on 05/03/2006. The European Medical Device Agency, the Italian Competent Authority and British Standards Institution were notified by email on 05/03/2006.
- Quantity in commerce
- 21 units
- Distribution
- AL, AR, FL, LA, MA, NJ, NY, PA, TN, TX, UT and Italy
- Product codes and lots
- Catalog number: 700-102846 , Lot number 25969