Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1140-06

Accu-Chek Ultraflex-8 infusion set; Catalog number 4544773001

Disetronic Medical Systems / initiated 2006-03-31 / Terminated

Disetronic Medical Systems, Inc. began a recall of Accu-Chek Ultraflex-8 infusion set; Catalog number 4544773001 on . The reason FDA records is "The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1140-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Disetronic Medical Systems
Product
Accu-Chek Ultraflex-8 infusion set; Catalog number 4544773001.
Product code
LZG Pump, Infusion, Insulin
Reason for recall
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
FDA root cause
Other
Action
"U.S. and International distributors, patients and their physicians were notified via letters dated 3/31/06. Distributors were instructed to search for all part numbers affected by the recall and to provide all affected customers and sub-distributors with the recall notice. Customers were offered a replacement set or replacement with a different model. A press release was also issued on 4/3/06.
Quantity in commerce
225,990 in US and Canada plus 6,750 outside North America
Distribution
Worldwide including USA, Argentina, Austria, Barbados, Belgium, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Honduras, Hungary, Italy, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
Product codes and lots
Lots 31097UF, 3C135UF, 3D001UF, 3E008UF, 3E065UF, 3G012UF, 3G107UF, 3G116UF, 3H040UF, 3H051UF, 3J018UF, 3K035UF, 3K101UF, 3L098UF, 3L122UF, 4A127UF, 4D030UF, 4D032UF, 4D161UF, 4D180UF, 4F020UF, 4F083UF, 4G184UF, 4H101IF, 4K133UF and 4L096UF.