SoftPath GUI Release Software Versions 3.17 and 4.1. Used in pathology laboratory for receiving accessing...
SCC Soft Computer began a recall of SoftPath GUI Release Software Versions 3.17 and 4.1. Used in pathology laboratory for receiving accessing specimen reports, billing and management reports for... on . The reason FDA records is "Loss of text misrepresented individual tissue diagnosis on the final diagnosis print out from the HIS system software. The specimen heading "J" was removed and the diagnosis appeared under the specimen heading "I" as a...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2329-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SCC Soft Computer
- Product
- SoftPath GUI Release Software Versions 3.17 and 4.1. Used in pathology laboratory for receiving accessing specimen reports, billing and management reports for surgical, GYN, NON-GYN and Autopsy cases.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- Loss of text misrepresented individual tissue diagnosis on the final diagnosis print out from the HIS system software. The specimen heading "J" was removed and the diagnosis appeared under the specimen heading "I" as a result the specimen with the heading "I" had two diagnosis.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Software corrections were sent as of 10/30/2003 for clients to upgrade to, the correction software patch is in versions 3.17.5.16, 3.17.6.10, 3.17.7.3 and 4.1.1.0. The firm did not inform the non-affected consignees of the failure because the firm concluded that the associated risk was low.
- Quantity in commerce
- 335
- Distribution
- Worldwide Distribution including USA and the country of Canada.
- Product codes and lots
- Versions 3.17 and 4.1