Z-1205-06
3D-I/III Ceiling Stand. Diagnostic x-ray tube mount
Siemens Medical Solutions USA, Inc. / initiated 2006-04-28 / Terminated
Siemens Medical Solutions USA, Inc began a recall of 3D-I/III Ceiling Stand. Diagnostic x-ray tube mount on . The reason FDA records is "Firm became aware that a potential pinch point can exist between the stop level and the safety catch when rotating the x-ray tube assembly. Risk of injury to operators fingers". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1205-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- 3D-I/III Ceiling Stand. Diagnostic x-ray tube mount
- Product code
- IYB Tube Mount, X-Ray, Diagnostic
- Reason for recall
- Firm became aware that a potential pinch point can exist between the stop level and the safety catch when rotating the x-ray tube assembly. Risk of injury to operators fingers.
- FDA root cause
- Other
- Action
- The recalling firm has issued a field correction per Update Instructions AX012/06/S. This update instructs a Siemens Customer Service Engineer to visit affected consignees to install a protective cover and additional warning labels on the device, beginning 04/28/2006.
- Quantity in commerce
- 466 units
- Distribution
- The products were shipped nationwide to medical facilities.
- Product codes and lots
- Model numbers 0923/G0923, 1022, 1080, 1212687, 1371970, 1372325, 1654250, 1576164, 8186751, 7388002, 7389489, 1738509, 1738517, 1738525, 7276504, 7276512, 7276520, 7276553, 7465008, 7486822, and 8185530.