Z-1174-06
RapidScreen RS-Digital, Medical Image Analyzer
Riverain Technologies / initiated 2006-05-04 / Terminated
Riverain Medical Group began a recall of RapidScreen RS-Digital, Medical Image Analyzer on . The reason FDA records is "The device has an actual sensitivity of 58.2% and a specificity of 3.9% (false positives per image). The product is approved to have a sensitivity of 63.3% and a specificity of 5.0. Thus, there is a difference of -5.1%...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1174-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Riverain Technologies
- Product
- RapidScreen RS-Digital, Medical Image Analyzer.
- Product code
- MYN Analyzer, Medical Image
- Reason for recall
- The device has an actual sensitivity of 58.2% and a specificity of 3.9% (false positives per image). The product is approved to have a sensitivity of 63.3% and a specificity of 5.0. Thus, there is a difference of -5.1% in sensitivity and +1.1 in specificity.
- FDA root cause
- Other
- Action
- The recalling firm sent a letter dated 5/4/06 to medical facilities, sent a letter dated 5/5/06 to distributors, and sent a letter dated 5/10/06 to GE Healthcare, Waukesha, WI.
- Quantity in commerce
- 72 devices
- Distribution
- Nationwide. The product was distributed to consignees located in IL, CA, MO, CO, WA, MT, ID, OR, NC, TX, TN, FL, UT, MN, OK, WI, OH, AZ, IA, and PA.
- Product codes and lots
- Serial numbers: 58HVH31, F8HVH31,49HVH31, 87HVH31, 47HVH31, F7HVH31, H7HVH31, 2RCTH31, 1RCTH31, B8HVH31, GQCTH31, G031M31, C031M31, 3131M31, 8031M31, 1131M31, 2FNNS31, BDNNS31,GDNNS31, FDNNS31, JDNNS31, GFNNS31, 2433X31, D333X31, F433X31, 5L3R241, 7L3R241, 148HN41, 648HN41, 348HN41, 848HN41, 838HN41, BR5VQ41, 4R5VQ41, JL0VQ41, JL0VQ41, 3N0VQ41, 2P0VQ41, BPZTQ41, FL0VQ41, 946VQ41, 296VQ41, 9M0VQ41, CL0VQ41, H86VQ41, 1QCWQ41, 1LCWQ41, 5LCWQ41, JNCWQ41, 596VQ41, 4NCWQ41, 9LCWQ41, GM0VQ41, BNCWQ41, 8QCWQ41, DN0VQ41, F26VQ41, B86VQ41, G96VQ41, 996VQ41, JR5VQ41, 2BCKM51, F9CKM51,79CKM51, H9CKM51, 39CKM51, C9CKM51, 1BCKM51, 59CKM51, 4BCKM51, 20-168, 20-160, and 7S5VQ41.