GYRUS ACMI PKS PlasmaSEAL Open Forceps, model 2103PK and 917015PK. Sterile EO. The device is intended to be...
Gyrus Medical, Inc. began a recall of GYRUS ACMI PKS PlasmaSEAL Open Forceps, model 2103PK and 917015PK. Sterile EO. The device is intended to be used with the PlasmaKinetic electrosurgical... on . The reason FDA records is "Gyrus ACMI has identified a product issue wherein their PlasmaSEAL Open Forceps, models 2103PK and 917015PK, may not deliver adequate hemostasis during use. Although Gyrus ACMI has had no reports of serious injury to...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1175-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Gyrus ACMI, Inc.
- Product
- GYRUS ACMI PKS PlasmaSEAL Open Forceps, model 2103PK and 917015PK. Sterile EO. The device is intended to be used with the PlasmaKinetic electrosurgical generator, software version 1.93 or greater or the PlasmaKinetic with SuperPulse electrosurgical generator, software version 2.19A or greater or the G400 electrosurgical generator. Made in USA Gyrus ACMI, Gyrus Medical, Inc., 6655 Wedgwood Road, Suite 160, Maple Grove, MN 55311-3602.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Gyrus ACMI has identified a product issue wherein their PlasmaSEAL Open Forceps, models 2103PK and 917015PK, may not deliver adequate hemostasis during use. Although Gyrus ACMI has had no reports of serious injury to patient, testing indicated that the possibility exists.
- FDA root cause
- Other
- Action
- US accounts were initially contacted by phone beginning 05/15/2006 and then sent a follow-up letter sent 05/31/2006. Customers were told of the issue, instructed to discontinue use of the product, told to remove all 2103PK and 917015PK devices from shelves and to quarantine product until they receive a letter with instructions to return the affected devices to Gyrus ACMI.
- Quantity in commerce
- 2971 (2931 domestic, 40 OUS)
- Distribution
- Nationwide, Australia, Canada, Netherlands. (AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, VA, WA, WI, WV).
- Product codes and lots
- Catalog No. 2103PK and 917015PK all production codes