Abbott brand FreeStyle Blood Glucose Meter, distributed by Abbott Diabetes Care, 1420 Harbor Bay Pwky...
Abbott Diabetes Care, Inc. began a recall of Abbott brand FreeStyle Blood Glucose Meter, distributed by Abbott Diabetes Care, 1420 Harbor Bay Pwky. Alameda, CA 94502. Manufactured by: Flextronics... on . The reason FDA records is "The products may encounter display problem, 'Er 4' message, during prolonged use when the low battery symbol is displayed. The situation can render the meter either inoperable or operable with invalid user configuration...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1306-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Abbott brand FreeStyle Blood Glucose Meter, distributed by Abbott Diabetes Care, 1420 Harbor Bay Pwky. Alameda, CA 94502. Manufactured by: Flextronics International, Long Jing Industrial Estate, Long Jing Road, Xi Li Town, Nantou District, Shenzhen, China
- Product code
- NBW System, Test, Blood Glucose, Over The Counter
- Reason for recall
- The products may encounter display problem, 'Er 4' message, during prolonged use when the low battery symbol is displayed. The situation can render the meter either inoperable or operable with invalid user configuration data including selectable unit of measure, and strip calibration code.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- On 5/22/06, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected products and providing instructions for the corrective actions.
- Quantity in commerce
- approx. 2,365,000 meters (both recall numbers)
- Distribution
- Worldwide
- Product codes and lots
- All codes.