Roche Calcium Reagent Kit for use on the Cobas MIRA Chemistry System; in vitro diagnostic; Cat. No...

Roche Diagnostics GmbH / initiated 2006-06-07 / Terminated / root cause: labeling and UDI

Roche Diagnostics Corp. began a recall of Roche Calcium Reagent Kit for use on the Cobas MIRA Chemistry System; in vitro diagnostic; Cat. No. 04585984160 on . The reason FDA records is "Falsely low patient and control results may be reported when the reagent has been on-board for less than the 8 hour use period specified in the labeling". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1176-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche Calcium Reagent Kit for use on the Cobas MIRA Chemistry System; in vitro diagnostic; Cat. No. 04585984160.
Product code
CIC Cresolphthalein Complexone, Calcium
Reason for recall
Falsely low patient and control results may be reported when the reagent has been on-board for less than the 8 hour use period specified in the labeling.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Consignees were notified via letter dated 6/6/06 and sent via UPS to discontinue use of these lots and to destroy them. Distributors are being asked to notify their customers.
Quantity in commerce
933 kits
Distribution
Nationwide.
Product codes and lots
Lots A157001 (exp. 7/31/07) and A157002 (exp. 9/30/07).