Z-1212-06

Product is Reprocessed Ethicon Bladeless Trocars. Model: 35NLT, 35NST, 35OL

Stryker Sustainability Solutions / initiated 2006-03-17 / Terminated / root cause: packaging

Vanguard Medical Concepts, Inc began a recall of Product is Reprocessed Ethicon Bladeless Trocars. Model: 35NLT, 35NST, 35OL on . The reason FDA records is "Review of complaints and testing show that the distal tip of the trocars may separate while in use". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1212-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Sustainability Solutions
Product
Product is Reprocessed Ethicon Bladeless Trocars. Model: 35NLT, 35NST, 35OL
Product code
NLM Laparoscope, General & Plastic Surgery, Reprocessed
Reason for recall
Review of complaints and testing show that the distal tip of the trocars may separate while in use.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
The firm sent recall notification letters to its consignees on 3-17-2006. They were instructed to discontinue use of the 35NLT, 35NST and 35Ol Reprocessed Optiview Trocars with expiration date of 12/20008 or earlier immediately. Any of this product in inventory were to be returned to the manufacturer for credit. The firm also offered the assistance of its representatives in returning all affected profucts. The investigation is ongoing, and corrective actions associated with the root cause of tips breaking off have been identified and are being implemented.
Quantity in commerce
13,842
Distribution
Nationwide.
Product codes and lots
All lot numbers produced before January, 1, 2006. (Product with expiration date of 12/20008 or earlier.)