Z-1161-06

OBSERVA R02 (version R02.00.17) computers using PSC 6000 or HHP Barcode Scanners which are connected to...

bioMerieux, Inc. / initiated 2006-03-28 / Terminated / root cause: software

bioMerieux Inc began a recall of OBSERVA R02 (version R02.00.17) computers using PSC 6000 or HHP Barcode Scanners which are connected to BacT/ALERT 3D on . The reason FDA records is "Software problem with scanner inter-character delay which may result in false positive or false negative results after incorrect scanning of bottle ID's". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1161-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
OBSERVA R02 (version R02.00.17) computers using PSC 6000 or HHP Barcode Scanners which are connected to BacT/ALERT 3D
Product code
MDB System, Blood Culturing
Reason for recall
Software problem with scanner inter-character delay which may result in false positive or false negative results after incorrect scanning of bottle ID's
Root cause tag
software
FDA root cause
Other
Action
The recalling firm telephoned customers who ordered the MB bottles and use the OBSERVA software ver. 2.0 to inform them of the issues. Recall letters were issued to all customers via overnight mail on 5/22/06 informing them a field representative would be visiting to make a short-term correction to the system. The letter also informed the customer a permanent correction to the software was scheduled for release later in the year.
Quantity in commerce
51 units
Distribution
WorldwideDistribution--Nationwide to medical centers. Government distribution was made to MN and CA, and military distribution was made to OK. Foreign distribution was made to Thailand, The Netherlands, Germany, Greece, Portugal, and France.
Product codes and lots
Software version R.02.00.17