VITALITY AVT, model A155. Implantable Cardioverter Defibrillator. Atrial and Ventricular Therapies. Guidant...
Boston Scientific CRM Corp began a recall of VITALITY AVT, model A155. Implantable Cardioverter Defibrillator. Atrial and Ventricular Therapies. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline... on . The reason FDA records is "Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may...". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-1291-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- VITALITY AVT, model A155. Implantable Cardioverter Defibrillator. Atrial and Ventricular Therapies. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.
- Product code
- LWP Implantable Pulse Generator, Pacemaker (Non-Crt)
- Reason for recall
- Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.
- Root cause tag
- supplier
- FDA root cause
- Other
- Action
- An Urgent Medical Device Safety Information & Corrective Action letter, dated May 12, 2006, was sent to physicians. The letter gave a description of the issue, clinical implications, rate of occurrence, devices affected and recommendation. Physicians are asked to follow-up with affected patients as soon as possible because ERI to EOL time may be shortened. Physicians are asked to contact Guidant for instructions on performing a baseline 'save to disk' and provide this information to Guidant. Guidant will be able to analyze the device memory data and estimate remaining longevity and provide individualized follow-up and replacement guidelines. Subsequent follow-up visits should continue to monitor for evidence of premature battery depletion.
- Quantity in commerce
- 66
- Distribution
- Nationwide including Puerto Rico. Worldwide, including Argentina, Belgium, Canada, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, New Zealand, Portugal, Spain, Syrian Arab Republic, United Arab Emirates
- Product codes and lots
- model A155, serial numbers: 113567, 113568, 113569, 113570, 113572, 113574, 113585, 113586, 113587, 113591, 113612, 113613, 113616, 113617, 113618, 113619, 113620, 113621, 113623, 113624, 113626, 113627, 113628, 113629, 113631, 113633, 113635, 113637, 113639, 113640, 113641, 113654, 113659, 113660, 113672, 113674, 113675, 113676, 113678, 113679, 113680, 113681, 113682, 113684, 113686, 113687, 113690, 113693, 113694, 113695, 113696, 113778, 113779, 113781, 113783, 113784, 113789, 113791, 113792, 113814, 113908, 113910, 113912, 113913, 113914, 114083