Z-1242-06

Kimberly-Clark/Ballard medical R2 triangle shaped non-radiolucent multifunction electrode products for adults

Kimberly-Clark Corporation / initiated 2006-05-18 / Terminated / root cause: design

Kimberly-Clark Corporation began a recall of Kimberly-Clark/Ballard medical R2 triangle shaped non-radiolucent multifunction electrode products for adults on . The reason FDA records is "The tinfoil in the triangle electrodes has the potential to develop cracks which may hinder electrical flow and product performance during defibrillation". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1242-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Kimberly-Clark Corporation
Product
Kimberly-Clark/Ballard medical R2 triangle shaped non-radiolucent multifunction electrode products for adults.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
The tinfoil in the triangle electrodes has the potential to develop cracks which may hinder electrical flow and product performance during defibrillation.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were first contacted by phone with follow up letter sent via fax or mail on May 18, 2006. International consignees will be notified through the firm's regional offices.
Quantity in commerce
95,639 electrode sets
Distribution
Worldwide ( including USA, Switzerland, Germany, UK, Canada, Australia, Netherlands, Italy, Greece, Spain, Hong Kong, Hungary, Denmark, Singapore, New Zealand, Ireland, Korea)
Product codes and lots
Catalog number 30-585, Lot number 300847; Catalog number 30-590 Lot numbers 28802, 297261, 304750; Catalog number 30-591, Lot number 299214; Catalog number 30-593, Lot number 288962, 293213, 294860, 302657; Catalog number 3100-1702, Lot number 297735; Catalog number 3100-703, Lot number 282936, 293214, 300849, 303000; Catalog number 3111-1721, Lot numbers 282181, 282938, 286332, 286339, 288964, 288970, 290318, 290905, 291535, 292660, 293217, 294206, 294862, 295600, 296270, 297263, 298032, 299985, 302659, 303003, 304091, 304753, 305509;  Catalog number 3112-1731, Lot numbers 282183, 282939, 285955, 286335, 286342, 290320, 290907, 291537, 292662, 293220, 294208, 294866, 295603, 296273, 299218, 299987, 300853, 304093, 305512; Catalog number 3114-1741, Lot numbers 282185, 284030, 290322, 291538, 292664, 295604, 297266, 300854, 304095; Catalog number 3200-1715, Lot numbers 282188, 282942, 284031, 288004, 288005, 288966, 288967, 288974, 288975, 290911, 291541, 292667, 293224, 294210 294870, 295607, 296276, 297269, 298038, 299221, 299991, 301444, 302665, 303008, 304097, 304757, 305515;  Catalog number 3250-1780, Lot numbers 292086, 297270, 301445, 302666.