Z-1249-06

Hamilton brand RAPHAEL Ventilator (Software Versio 2.2x), Product is distributed by Hamilton Medical, Inc...

Hamilton Medical AG / initiated 2006-05-24 / Terminated / root cause: manufacturing process

Hamilton Medical Inc began a recall of Hamilton brand RAPHAEL Ventilator (Software Versio 2.2x), Product is distributed by Hamilton Medical, Inc., 4990 Energy Way,Reno, NV 89502-Serial Numbers 2975... on . The reason FDA records is "ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1249-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hamilton Medical AG
Product
Hamilton brand RAPHAEL Ventilator (Software Versio 2.2x), Product is distributed by Hamilton Medical, Inc., 4990 Energy Way,Reno, NV 89502-Serial Numbers 2975 to 5360,
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.
Root cause tag
manufacturing process
FDA root cause
Other
Action
On May 24, 2006 the manufacturer initiated this recall. The Recalling establishment/US Initial Importer/Distributor-received the manufacturer''s notice on May 31, 2006 and initiated their recall on June 1, 2006. The US distributor has issued a Medical Device Field Correction.
Quantity in commerce
2,385 for all models
Distribution
Worldwide
Product codes and lots
Serial Numbers 2975 to 5360