Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1203-06

ImmunoCard toxins A & B kit, a rapid enzyme immunoassay for the detection of Clostridium difficile Toxins A...

Meridian Bioscience Inc / initiated 2006-06-02 / Terminated

Meridian Bioscience Inc began a recall of ImmunoCard toxins A & B kit, a rapid enzyme immunoassay for the detection of Clostridium difficile Toxins A and B in Stool Samples (Catalog number 712050). An... on . The reason FDA records is "The product may produce non-specific reactions with negative specimens and/or the negative control reagent. These non-specific interactions may result in increased invalid test rates and/or positivity rates". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1203-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Meridian Bioscience Inc
Product
ImmunoCard toxins A & B kit, a rapid enzyme immunoassay for the detection of Clostridium difficile Toxins A and B in Stool Samples (Catalog number 712050). An in vitro diagnostic.
Product code
LLH Reagents, Clostridium Difficile Toxin
Reason for recall
The product may produce non-specific reactions with negative specimens and/or the negative control reagent. These non-specific interactions may result in increased invalid test rates and/or positivity rates.
FDA root cause
Other
Action
The recalling firm sent a recall letter dated 6/2/06 to their customers.
Quantity in commerce
704
Distribution
Nationwide.
Product codes and lots
Lot 712050.066, expiration date 2007-02-22 and Lot 712050.068, expiration date 2007-03-08.