Z-1313-06

Access Immunoassay system CEA, in vitro diagnostic, Part No. 33200

Siemens Healthcare Diagnostics Inc. / initiated 2006-04-12 / Terminated

Beckman Coulter Inc began a recall of Access Immunoassay system CEA, in vitro diagnostic, Part No. 33200 on . The reason FDA records is "Firm has confirmed that the Access CEA Assay may generate decreased values with certain frozen platelet specimens". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1313-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Immunoassay system CEA, in vitro diagnostic, Part No. 33200
Product code
DHX System, Test, Carcinoembryonic Antigen
Reason for recall
Firm has confirmed that the Access CEA Assay may generate decreased values with certain frozen platelet specimens.
FDA root cause
Other
Action
A Product Corrective Action letter was sent via US mail the week of April 10, 2006 to all Access CEA customers informing them that the Access CEA Assay could generate decreased values with certain frozen patient specimens. Although they have found it at a low occurrence, it is recommended that specimens should not be frozen prior to analysis until further notice. The letter informs customers that Beckman Coulter has verified sample stability through at least 14 days when properly stored at 2-8 degrees Celsius and shipping stability for 4 days at 2-8 degrees Celsius using cool packs. Firm also advises customers to contact customer technical support or local Beckman Coulter Representative if laboratory is unable to avoid freezing specimens and to review results from frozen CEA samples to ensure that the match the total clinical presentation per the customer''s established laboratory procedure. The letter requests that the notification remain part of the laboratory Quality System documentation and that the response form be returned to the firm within 10 days.
Quantity in commerce
775
Distribution
Nationwide and Canada
Product codes and lots
All codes