Z-0001-2007

GE 4 inch Anterior Cervical Post, GE Part Number 1006385 or 1006385-NAV, GE Healthcare OEC Medical Systems...

GE Healthcare / initiated 2006-05-31 / Terminated / root cause: manufacturing process

GE OEC Medical Systems, Inc began a recall of GE 4 inch Anterior Cervical Post, GE Part Number 1006385 or 1006385-NAV, GE Healthcare OEC Medical Systems (an accessory used with the InstaTrak 3500 plus and... on . The reason FDA records is "Weld defect- the weld between the post body and the screw was not properly formed and can fail, even when minimal force is used". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0001-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE 4 inch Anterior Cervical Post, GE Part Number 1006385 or 1006385-NAV, GE Healthcare OEC Medical Systems (an accessory used with the InstaTrak 3500 plus and 9800 C-Arm Navigation systems)
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
Weld defect- the weld between the post body and the screw was not properly formed and can fail, even when minimal force is used.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified by letter on 05/31/2006
Quantity in commerce
143 units
Distribution
Nationwide, including one VA facility in PA. No military distribution. Foreign distribution to Canada, China, Italy, Japan, Spain.
Product codes and lots
ALL LOT NUMBERS -GE Part number 1006385 or 1006385-NAV