Z-1314-06
D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer
Radiometer America, Inc / initiated 2005-02-01 / Terminated / root cause: manufacturing process
Radiometer America Inc began a recall of D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer on . The reason FDA records is "The CI membrane units generate too low values for status and sensitivity during calibration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1314-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Radiometer America, Inc
- Product
- D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer.
- Product code
- CGZ Electrode, Ion-Specific, Chloride
- Reason for recall
- The CI membrane units generate too low values for status and sensitivity during calibration.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- The recalling firm sent a letter, dated 2/05, to all of their consignees.The letter instructed the customers to discontinue use of the product and discard any remaining stock which they had on hand. The letter was accompanied by replacement chloride membranes to replace the customer''s portion of the suspect lot.
- Quantity in commerce
- 85/4 unit boxes
- Distribution
- Nationwide.
- Product codes and lots
- Lot Code: R0053.