Z-1314-06

D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer

Radiometer America, Inc / initiated 2005-02-01 / Terminated / root cause: manufacturing process

Radiometer America Inc began a recall of D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer on . The reason FDA records is "The CI membrane units generate too low values for status and sensitivity during calibration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1314-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Radiometer America, Inc
Product
D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer.
Product code
CGZ Electrode, Ion-Specific, Chloride
Reason for recall
The CI membrane units generate too low values for status and sensitivity during calibration.
Root cause tag
manufacturing process
FDA root cause
Other
Action
The recalling firm sent a letter, dated 2/05, to all of their consignees.The letter instructed the customers to discontinue use of the product and discard any remaining stock which they had on hand. The letter was accompanied by replacement chloride membranes to replace the customer''s portion of the suspect lot.
Quantity in commerce
85/4 unit boxes
Distribution
Nationwide.
Product codes and lots
Lot Code: R0053.