Z-1315-06
Medplus ER Drug Screen Test Model #5800KAB packaged as 25 individually pouched devices in a labeled kit box
Applied Biotech Inc / initiated 2005-05-31 / Terminated
Applied Biotech Inc began a recall of Medplus ER Drug Screen Test Model #5800KAB packaged as 25 individually pouched devices in a labeled kit box on . The reason FDA records is "The firm has discovered that a number of distributed devices of the Medplus ER Drug Screen Test may be incorrectly assembled which may give incorrect results. The product is an in vitro test for rapid detection of...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1315-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Applied Biotech Inc
- Product
- Medplus ER Drug Screen Test Model #5800KAB packaged as 25 individually pouched devices in a labeled kit box.
- Product code
- LCM Enzyme Immunoassay, Phencyclidine
- Reason for recall
- The firm has discovered that a number of distributed devices of the Medplus ER Drug Screen Test may be incorrectly assembled which may give incorrect results. The product is an in vitro test for rapid detection of various drugs such as Amphetamines, Barbiturates, Phencyclidine, Cocaine, Methamphetamine, Tricyclic Antidepressants, Morphine, Tetrahydrocannabinol, & Benzodiazepines in human urine.
- FDA root cause
- Other
- Action
- Recall notifications that describe the problem, dated May 31,2005, were sent via fax. Customers were instructed to block all stock of affected lots in their warehouse, contact their primary consignees and advise them to discontinue use immediately and dispose any unused materials. A faxback form was used to communicate to the recalling firm the quantity blocked and disposed of. Arrangements were made for providing replacement product.
- Quantity in commerce
- Lot #85699: 71 kits (1775 devices) & Lot #85671: 44 kits (1100 devices)
- Distribution
- Nationwide (distributors in KS and OR).
- Product codes and lots
- Lot Numbers Affected: 85699 and 85671