Z-1184-06
Theratron Elite 100 Cobalt Radiotherapy
Bwxt Medical Ltd. / initiated 2004-07-08 / Terminated
MDS Canada Inc. DBA MDS Nordion began a recall of Theratron Elite 100 Cobalt Radiotherapy on . The reason FDA records is "The source locking device may not fit properly through the smaller opening in the new field light shield which could prevent the source from being returned manually under emergency conditions". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1184-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bwxt Medical Ltd.
- Product
- Theratron Elite 100 Cobalt Radiotherapy
- Product code
- IWB System, Radiation Therapy, Radionuclide
- Reason for recall
- The source locking device may not fit properly through the smaller opening in the new field light shield which could prevent the source from being returned manually under emergency conditions.
- FDA root cause
- Other
- Action
- A User Bulletin (CUB-2004-01dated 7/5/04) was issued to affected users on 8/24/04 and 8/31/04. The bulletin provided 2 options for corrective action.
- Quantity in commerce
- 17 units
- Distribution
- The products were shipped to medical facilities in FL and KS. The product was also shipped to Kuwait, Spain, Honduras, Romania, France, Grrece, Bangladesh, Germany, Nepal, Canada, England, Thailand, Phillippines, Mexico, Columbia, India, Switzerland, Syria, Trinidad & Tobago, Ireland, Turkey, Algeria, Portugal, Poland, Iran, China, Cuba, Belgium, Hungary, Egypt, Italy, Vietnam, Sri Lanka, and Costa Rica.
- Product codes and lots
- Model Theratron Elite 100