Z-1406-06

D7120, BG-OX Cartridge, D7120 REF 946-003, and 946-001, cuvettes for the NPT7 series of blood gas analyzers

Radiometer America, Inc / initiated 2005-01-12 / Terminated / root cause: manufacturing process

Radiometer America Inc began a recall of D7120, BG-OX Cartridge, D7120 REF 946-003, and 946-001, cuvettes for the NPT7 series of blood gas analyzers on . The reason FDA records is "QC Bias-a solution used in the manufacturing of the membrane had a different composition which introduces a bias of 20 mpH". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1406-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Radiometer America, Inc
Product
D7120, BG-OX Cartridge, D7120 REF 946-003, and 946-001, cuvettes for the NPT7 series of blood gas analyzers.
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
QC Bias-a solution used in the manufacturing of the membrane had a different composition which introduces a bias of 20 mpH.
Root cause tag
manufacturing process
FDA root cause
Other
Action
The firm contacted their customers and informed them of this problem and provided a workaround procedure, in a a letter dated 1/12/2005. A ''Field Action Notification'' (FAN) was issued to the firm''s Field Service personnel on 1/21/2005. The letter instructs their personnel to update the NPT7 application software on all units to V3.05 and to hand over a copy of the letter ''Changes related to the Qualicheck2+ and NPT7,'' to the user/customers. In addition, the firm''s personnel were instructed to remove any of the suspect cartridges found on-hand at each customer location; destroy them;and replace them with different lot numberd product.
Quantity in commerce
545
Distribution
The recalled device component was distributed to a total of 56 customers, nationwide located in the following states: AL, AR, AZ, CA, CO, GA, ID, IL, KA, MA, ME, MI, MO, MS, NC, ND, NE, NH, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT & WA.
Product codes and lots
Lot R0187, Expiration Date: Sept. 2005; and Lot R0016, Expiration Date: 08/2005.