Z-1319-06

D711, Reference Membranes, Part #942-058, used with the ABL835 Blood Gas Analyzers

Radiometer America, Inc / initiated 2005-06-17 / Terminated

Radiometer America Inc began a recall of D711, Reference Membranes, Part #942-058, used with the ABL835 Blood Gas Analyzers on . The reason FDA records is "The recalled membranes cause the status value of the pH and electrolyte electrodes to be offset causing readings outside of the acceptable range". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1319-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Radiometer America, Inc
Product
D711, Reference Membranes, Part #942-058, used with the ABL835 Blood Gas Analyzers.
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
The recalled membranes cause the status value of the pH and electrolyte electrodes to be offset causing readings outside of the acceptable range.
FDA root cause
Other
Action
The recalling firm notified their distributor/customers of this recall, via telephone on 6/17/2005, at which time, the customers were asked to destroy any and all suspect reference membranes found on-site and at their subaccount customer locations.
Quantity in commerce
13 boxes/4 membranes per box
Distribution
Nationwide, including TX, OK, MI, CO, NY, and Canada.
Product codes and lots
Lot Code: R0308, Expiration Date: 4/07; Lot Code: R0309, Expiration Date: 5/07; and Lot Code: R0310, Expiration Date: 5/07.