Z-1453-06

TCM4 Series Monitoring System (Base Unit) 391-876 (affected device) CompactFlash cards: 914-698 (Defective...

Radiometer America, Inc / initiated 2006-04-07 / Terminated

Radiometer America Inc began a recall of TCM4 Series Monitoring System (Base Unit) 391-876 (affected device) CompactFlash cards: 914-698 (Defective Device Component) on . The reason FDA records is "System shut down-When the TCM Monitor is turned on and the booting process begins, the device stops after the memory count and will not proceed further". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1453-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Radiometer America, Inc
Product
TCM4 Series Monitoring System (Base Unit) 391-876 (affected device) CompactFlash cards: 914-698 (Defective Device Component)
Product code
KLK Monitor, Oxygen, Cutaneous, For Infant Not Under Gas Anesthesia
Reason for recall
System shut down-When the TCM Monitor is turned on and the booting process begins, the device stops after the memory count and will not proceed further.
FDA root cause
Other
Action
The firm began a recall/corrective action in response to this component defect on 4/7/2006, at which time, the firm issued a Field Action Notes memo (915-234). The firm is having their representatives: 1) assure that the use of the CompactFlashcards manufactured by Simple Tech has been discontinued at each customer location; 2) Replaace the CompactFlash card in all basic Units, 391-876, of R0036 through R0043; and 3) replace any spare part #914-698, which has been installed in Basic Units, 391--876, below R0036. Once completed, the firm has supplied their customers with a TELEFAX form to be completed, signed and dated by each customerand faxed back to Radiometer Medical ApS. The form documents that the instructions provided in theField Action Notes have been followed and that the necessary component replacements have been completed. The firm anticipates that the necesary corrective action will be completed by 7/1/2006.
Quantity in commerce
61 in U.S. 7 in Canada
Distribution
Worldwide-The recalled product was distributed to customers throughout the U.S. and Canada.
Product codes and lots
TCM4 - Base Unit : Model: 391-876, Lot Numbrs R0036 thru R0043 (both inclusive). CompactFlash cards - Component: Model #914-698