Z-1437-06
TC Monitor - Transcutaneous Monitor for p02 and pC02, Model #TCM4, (Cutaneous Oxygen Monitor)
Radiometer America, Inc / initiated 2005-02-02 / Terminated
Radiometer America Inc began a recall of TC Monitor - Transcutaneous Monitor for p02 and pC02, Model #TCM4, (Cutaneous Oxygen Monitor) on . The reason FDA records is "Alarm may be silenced- Rapid switching, on and off, of the TCM4 monitor may cause the built-in acoustical alarm to remain silent when tripped". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1437-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Radiometer America, Inc
- Product
- TC Monitor - Transcutaneous Monitor for p02 and pC02, Model #TCM4, (Cutaneous Oxygen Monitor)
- Product code
- KLK Monitor, Oxygen, Cutaneous, For Infant Not Under Gas Anesthesia
- Reason for recall
- Alarm may be silenced- Rapid switching, on and off, of the TCM4 monitor may cause the built-in acoustical alarm to remain silent when tripped.
- FDA root cause
- Other
- Action
- The firm initiated a recall action for the suspect device with a letter dated 2/2/2005. During a site visit to the U.S. customers conducted by the firm''s Field Service reps, the reps informed the user/customers of this problem and requested that all use of the device be discontinued, pending the formulation of a more permanent solution. Beginning on 2/8/2005, a second site visit was made to the same user/customers, at which time, the affected units were upgraded with operating software version 2.12. The firm''s recall/corrective action was scheduled for completion by 3/1/2005.
- Quantity in commerce
- 9 monitors were distributed.
- Distribution
- Nationwide
- Product codes and lots