Z-0079-2007

CryoCath Frost Byte CryoSurgical Clamp, Model 60FB1, Manufacturer CryoCath Technologies, Inc., Quebec...

Medtronic, Inc. (Minneapolis, MN 55447) / initiated 2006-05-05 / Terminated / root cause: labeling and UDI

ATS Medical, Inc. began a recall of CryoCath Frost Byte CryoSurgical Clamp, Model 60FB1, Manufacturer CryoCath Technologies, Inc., Quebec, Canada. The product consists of a cryosurgical probe... on . The reason FDA records is "Incorrect Expiration Date: Three lots of the FrostByte (TM) clamp packaged with SurgiFrost 7 cm cryosurgical probes have incorrect expiration dates. These lots are labeled with a two (2) year expiratation date while the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0079-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc. (Minneapolis, MN 55447)
Product
CryoCath Frost Byte CryoSurgical Clamp, Model 60FB1, Manufacturer CryoCath Technologies, Inc., Quebec, Canada. The product consists of a cryosurgical probe (SurgiFrost 7 cm) plus Clamp (FrostByte) packaged for use with Cryosurgical console.
Product code
GEH Unit, Cryosurgical, Accessories
Reason for recall
Incorrect Expiration Date: Three lots of the FrostByte (TM) clamp packaged with SurgiFrost 7 cm cryosurgical probes have incorrect expiration dates. These lots are labeled with a two (2) year expiratation date while the correct date is one (1) year.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Consignee''s were notified by ATS Medical Representatives via visits. Product was either relabeled with correct expiration date or retrieved by ATS Medical representatives.
Quantity in commerce
219 units
Distribution
Nationwide-ATS Medical distirubted devices to hospitals located in the following states IN, KS, MO, TX. There may be other US distributors of this product.
Product codes and lots
Lot: 50515, 50602, 50609